Drug dissolution test as a quality control test: In simple terms, at present a dissolution test as a QC test means conducting a dissolution test as described in a pharmacopeia, in particular USP. If the test meets the pharmacopeial (Tolerances) requirements then the product may be considered a “Quality” product. It is, however, not clear what “quality” the test refers to or to what product property the test is linked to? Therefore, to overcome this lack of objectivity/relevancy, pharmacopeial tests are described as “consistency” tests. Again, it is not clear the “consistency” of which property or parameter the test is refers to?

Dissolution testing during product development: One of the main uses of dissolution testing is to facilitate product development. This use is based on the principle that the tests should be able to provide potential in vivo drug release behavior information. However, this is a commonly recognized fact that currently used dissolution tests generally do not provide in vivo relevant product characteristics. Thus, current practices of dissolution testing at the product development stage appear to lack relevancy and objectivity.    

Practices of methods development: For developing a method one requires a well established and accepted reference (product or parameter). In this case, a reference product should be available with known drug dissolution results which are established independently. As there is no reference product with known dissolution results available at present, it is not possible to develop a dissolution method. Developing a method, as commonly suggested in literature, based on test products (with unknown dissolution characteristics) is neither scientifically valid nor possible.

Dissolution apparatus selection: Selecting a dissolution apparatus should also be part of a method development exercise. An apparatus must be able to provide a relevant testing environment. Current practices appear to suggest that relevant apparatuses, and associated experimental conditions, are those which are described in the pharmacopeias only. On the other hand, it has never been shown that suggested apparatuses and the associated experimental conditions are suitable for their intended purpose. That is, the suggested apparatuses have never been validated for the intended purpose.

Selecting experimental conditions: Choices of experimental conditions such as rpm, buffers, described in the literatures appears to be arbitrary and random. Moreover, experimental conditions are selected to achieve some desired dissolution characteristics of the test products. Thus, a formulator/analyst would never know the “true” dissolution characteristics of the product. Experimental conditions must be linked to the physiological environment and be product independent. Drug and/or product dependent experimental conditions, as commonly described, are neither scientifically valid nor relevant.

Establishing IVIVC: Developing IVIVC commonly refers to RELATING in vitro dissolution results to in vivo dissolution results using a mathematical approach of de-convolution. However, dissolution tests in general are conducted based on the principle that in vitro release should relate well with in vivo results. Therefore, it is not clear why every manufacturer/analyst has to go through this exercise of establishing this in vitro-in vivo link for each and every product. In addition, as stated above that the current apparatuses/practices of dissolution testing have never been validated for providing relevant in vivo dissolution characteristics, then why one should be expected to achieve appropriate IVIVC?